ADS-300 Anesthesia workstation with AVS-360 Ventilator + 2 Gas

ADS-300 configured with AVS-360 anesthesia ventilator and a 2-gas 4-tube high-accuracy rotameter configuration. Integrated with the ACS-3 electric-driven circle absorber, the system requires no driving gas.

 

AVS-360 is an advanced, user-friendly, microprocessor-based anesthesia ventilator suitable for adult and pediatric patients. AVS-360 has a 7” TFT touch screen display with volume and pressure graphs. AVS-360 can be operated in VCMV, V-SIMV, PCMV, P-SIMV, PRVC, PSV, BAG, and Standby modes.

PARAMETERSAVS-360
Mode of OperationVCMV, V- SIMV, PCMV, P-SIMV, PRVC, PSV, BAG and Standby.
Patient TypeAdult and Child [Selectable]
I: E Ratio4:1 to 1:4
Plateau0 to 50%
Trigger0.5 to 5.0 cmH2O
Tidal Volume50 to 1500 ml.
Respiratory Rate1 to 60 Bpm
Ti0.2 to 8.0 Sec
Pi5 to 60 cmH2O
PEEP1 to 30 cmH2O
Pressure Support5 to 40 cmH2O
Ti max0.2 to 8.0 Sec
Backup RR6 to 60 Bpm
Backup ModeVCMV, V- SIMV, PCMV, P-SIMV and PRVC.
Monitoring ParametersTidal Volume, Minute Volume, RR, FIO2, Pi, P Mean, P Plat, PEEP Obtained.
AlarmsLow and High for Pressure, RR, Minute Volume and FIO2, Active patient, Patient disconnected, Driving Gas failure, low battery indication, Apnea.
Loop TypesFlow vs Pressure, Flow vs Volume & Volume vs Pressure [Selectable Any Two]
Graph TypesFlow , Pressure & Volume[Selectable Any Two]
Fresh Gas DecouplingEnabled
Flow SensorInternal & Permanent
Display7″ TFT with touch Screen.
Power Input120W
Driving GasNot Required (Electrical Bellow)
Gas Input Pressure3~6 Bar (Pipeline) / 4-150Bar (Cylinder)
Battery Backup2 hours (Lithium-ion)
Flow Meter Type2 Gas (4 Tubes)