ADS-300 Anesthesia workstation with AVS-180 Ventilator + 3 Gas

ADS-300 configured with AVS-180 anesthesia ventilator and a 3-gas 5-tube high-accuracy rotameter configuration. Integrated with the ACS-1 gas-driven circle absorber using air as the driving gas.

AVS-180 is an advanced, user-friendly, microprocessor-based anesthesia ventilator suitable for adult and pediatric patients. AVS-180 has a 7” TFT touch screen display with volume and pressure graphs. AVS-180 can be operated in CMV, ACMV, SIMV, PLV, BAG and Standby modes

PARAMETERS AVS-180
Mode of Operation CMV, ACMV, SIMV, PLV, BAG & Standby.
Patient Type Adult and Child [Selectable]
I: E Ratio 4:1 to 1:4
Plateau 0 to 50%
Trigger 0.5 to 5.0 cmH2O
Tidal Volume 50 to 1500 ml.
Respiratory Rate 1 to 40 Bpm
Ti 0.3 to 8.0 Sec
Pi 5 to 60 cmH2O
PEEP
Pressure Support
Ti max
Backup RR
Backup Mode CMV
Monitoring Parameters Tidal Volume, Minute Volume, RR, FIO2, Pi, P Mean, P Plat, PEEP Obtained.
Alarms Pressure Low and High, RR Low and High, FIO2 Low and High, Active patient, Patient disconnected, Driving gas failure, Low battery indication, Minute Volume low-High.
Loop Types Flow vs Pressure, Flow vs Volume & Volume vs Pressure [Selectable Any Two]
Graph Types Flow , Pressure & Volume[Selectable Any Two]
Fresh Gas Decoupling Enabled
Flow Sensor External
Display 7″ TFT with touch Screen.
Power Input 20 W
Driving Gas Air
Gas Input Pressure 3~6 Bar (Pipeline) / 4-150Bar (Cylinder)
Battery Backup 2 hours (Lithium-ion)
Flow Meter Type 3 Gas(5 Tubes)